User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Simpson 260 VOM /5100,3458 DATE: 19-Dec-95 AUTHOR: User Contributed REVISION: ADJUSTMENT THRESHOLD: 100% NUMBER OF TESTS: 30 NUMBER OF LINES: 214 CONFIGURATION: Fluke 5100B CONFIGURATION: HP 3458A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON # Equipment # Equip. Type: Simpson 260 VOM # Notes: Current to 1.9 Amps only. # Units with 5000V Ranges are calibrated to 1000V only. # # Review Approval:_____________________________ Release Date:____/___/____ # Revision Description. # 1.001 RSLT =Accuracy: Manufacturers specification except for the 1.002 RSLT = following: Current limited to 1.9 amps 1.003 RSLT = Resistance tolerance estimated @ 1.004 RSLT = +8% -7.5% of nominal. Based on 1.005 RSLT = MFR spec of ñ3ø of ARC. 1.006 RSLT = Units with a 5000V Range are calibrated to 1.007 RSLT = 1000V only. 1.008 RSLT =Reference Documents: Manufacturers manual. 1.009 RSLT = 1.010 ASK+ X 1.011 ASK- R N P F W 1.012 DISP Lay the Simpson on its back for this calibration. 1.013 DISP Zero the Meter Movement using the Center Screw. 1.014 OPBR Does the UUT have 250MV & +1V Input Terminals? 1.015 JMPF 3.003 1.016 HEAD {Volts DC and Linearity Tests} 1.017 DISP Connect the Red Test Lead to the 250mV Input Terminal 1.017 DISP of the Simpson. Connect the Black Test Lead to the 1.017 DISP COMMON - Terminal of the Simpson. 1.017 DISP 1.017 DISP Connect the other end of the Test Leads to the 1.017 DISP 'Binding Posts to Double Banana Plug Adapter' 1.017 DISP Red to Red and Black to Black. 1.018 DISP Set the Function Selector switch on the left hand side 1.018 DISP of the UUT to + D.C. 1.019 DISP Set the UUT Range to the 2.5V Range. 1.020 5100 250 250mV 2/ 2W #! Test Tol 0.005, Sys Tol 3.5e-005, TUR 142.857 (>= 4.00). 2.001 DISP Move the Red Test Lead from the 250mV input terminal 2.001 DISP to the +1V input terminal above the function switch. 2.002 5100 1 1V 2/ 2W #! Test Tol 0.02, Sys Tol 6.5e-005, TUR 307.692 (>= 4.00). 3.001 DISP Move the RED Test Lead from the +1V input terminal to 3.001 DISP the + terminal near the bottom left hand corner. 3.002 JMP 4.002 3.003 DISP Connect the Red Test Lead to the + Input Terminal 3.003 DISP of the Simpson. Connect the Black Test Lead to the 3.003 DISP COMMON - Terminal of the Simpson. 3.003 DISP 3.003 DISP Connect the other end of the Test Leads to the 3.003 DISP 'Binding Posts to Double Banana Plug Adapter' 3.003 DISP Red to Red and Black to Black. 3.004 DISP Set the Function Selector switch on the left hand side 3.004 DISP of the UUT to D.C. 3.005 DISP Set the UUT Range to the 2.5V Range. 3.006 5100 2.5 2.5V 2/ 2W #! Test Tol 0.05, Sys Tol 0.000305, TUR 163.934 (>= 4.00). 4.001 JMP 6.001 4.002 5100 2.5 2.5V 2/ 2W #! Test Tol 0.05, Sys Tol 0.000305, TUR 163.934 (>= 4.00). 5.001 DISP Set the UUT Function switch to - D.C. 5.002 5100 2.5 -2.5V 2/ 2W #! Test Tol 0.05, Sys Tol 0.000305, TUR 163.934 (>= 4.00). 6.001 DISP Set the UUT Function switch to + D.C. 6.001 DISP Set the UUT Range to the 10V Range. 6.002 5100 10 1V 2/ 2W #! Test Tol 0.2, Sys Tol 6.5e-005, TUR 3076.92 (>= 4.00). 7.001 5100 10 5V 2/ 2W #! Test Tol 0.2, Sys Tol 0.000405, TUR 493.827 (>= 4.00). 8.001 5100 10 10V 2/ 2W #! Test Tol 0.2, Sys Tol 0.000605, TUR 330.579 (>= 4.00). 9.001 DISP Set the UUT Range to the 50V Range. 9.002 5100 50 50V 2/ 2W #! Test Tol 1, Sys Tol 0.004, TUR 250.000 (>= 4.00). 10.001 DISP Set the UUT Range to the 250V Range. 10.002 5100 250 250V 2/ 2W #! Test Tol 5, Sys Tol 0.021, TUR 238.095 (>= 4.00). 11.001 OPBR Does the UUT have a 500V Range? 11.002 JMPF 12.003 11.003 DISP Set the UUT Range to the 500V Range. 11.004 5100 500 500V 2/ 2W #! Test Tol 10, Sys Tol 0.031, TUR 322.581 (>= 4.00). 12.001 DISP Move the RED test lead to the 1000V input terminal 12.001 DISP at the lower right of the Simpson. 12.002 JMP 12.004 12.003 DISP Set the UUT Range to the 1000V Range. 12.004 5100 1000 1000V 2/ 2W #! Test Tol 20, Sys Tol 0.051, TUR 392.157 (>= 4.00). 13.001 HEAD {Volts AC Tests} 13.002 DISP Set the UUT Function switch to A.C. 13.003 5100 1000 1000V 3/ 400H 2W #! Test Tol 30, Sys Tol 0.405, TUR 74.0741 (>= 4.00). 14.001 OPBR Does the UUT have a 500V Range? 14.002 JMPF 15.001 14.003 DISP Move the RED test lead from the 1000V input terminal 14.003 DISP to the + terminal near the bottom left corner of 14.003 DISP the meter. 14.004 5100 500 500V 3/ 400H 2W #! Test Tol 15, Sys Tol 0.23, TUR 65.2174 (>= 4.00). 15.001 DISP Set the UUT Range to the 250V Range. 15.002 5100 250 250V 3/ 400H 2W #! Test Tol 7.5, Sys Tol 0.1425, TUR 52.6316 (>= 4.00). 16.001 DISP Set the UUT Range to the 50V Range. 16.002 5100 50 50V 3/ 400H 2W #! Test Tol 1.5, Sys Tol 0.0275, TUR 54.5455 (>= 4.00). 17.001 DISP Set the UUT Range to the 10V Range. 17.002 5100 10 10V 3/ 400H 2W #! Test Tol 0.3, Sys Tol 0.00455, TUR 65.9341 (>= 4.00). 18.001 DISP Set the UUT Range to the 2.5V Range. 18.002 5100 2.5 2.5V 3/ 400H 2W #! Test Tol 0.075, Sys Tol 0.001925, TUR 38.9610 (>= 4.00). 19.001 HEAD {Frequency Response Tests} 19.002 5100 2.5 2.5V 3/ 200H 2W #! Test Tol 0.075, Sys Tol 0.001925, TUR 38.9610 (>= 4.00). 20.001 5100 2.5 2.5V 3/ 60H 2W #! Test Tol 0.075, Sys Tol 0.001925, TUR 38.9610 (>= 4.00). 21.001 HEAD {Current DC Tests} 21.002 DISP Set the UUT Function switch to +D.C. 21.003 OPBR Does the UUT have a 1 MA Range? 21.004 JMPF 22.001 21.005 DISP Set the UUT Range to the 1 MA Range 21.006 5100 1 1mA 2/ 2W #! Test Tol 2e-005, Sys Tol 2e-007, TUR 100.000 (>= 4.00). 22.001 DISP Set the UUT Range to the 10 MA Range. 22.002 5100 10 10mA 2/ 2W #! Test Tol 0.0002, Sys Tol 1.91e-006, TUR 104.712 (>= 4.00). 23.001 DISP Set the UUT Range to the 100 MA Range. 23.002 5100 100 100mA 2/ 2W #! Test Tol 0.002, Sys Tol 1.901e-005, TUR 105.208 (>= 4.00). 24.001 DISP Set the UUt Range to the 500 MA Range. 24.002 5100 500 500mA 2/ 2W #! Test Tol 0.01, Sys Tol 0.00011501, TUR 86.9490 (>= 4.00). 25.001 OPBR Does the UUT have a 50 æAMPS Range? 25.002 JMPF 26.002 25.003 DISP Set the UUT Range to 50V (u amps) 25.004 DISP Move the Red wire from the UUT + Input 25.004 DISP to the 50 æAMPS Input. 25.005 5100 50 50uA 1.5/ 2W #! Test Tol 7.5e-007, Sys Tol 2.15e-008, TUR 34.8837 (>= 4.00). 26.001 JMP 27.001 26.002 DISP Set the UUT Range to the 100 æAMPS Range. 26.003 5100 100 100uA 1.5/ 2W #! Test Tol 1.5e-006, Sys Tol 2.9e-008, TUR 51.7241 (>= 4.00). 27.001 DISP Set the UUT Range to 10mA (amps) 27.002 DISP Move the Red wire from the 50æAMPS or + Input 27.002 DISP Terminal to the +10A Input and the Black lead to the 27.002 DISP -10A Input of the meter. 27.003 5100 10 1.9A 2/ 2W #! Test Tol 0.2, Sys Tol 0.00032501, TUR 615.366 (>= 4.00). 28.001 HEAD {Resistance Tests} 28.002 DISP Move the Red wire from the UUTs +10A Input 28.002 DISP to the + Input near the bottom left hand corner. 28.002 DISP Move the Black wire from the -10A Input to the 28.002 DISP - COMMON Input Terminal of the Simpson. 28.003 DISP Set the Range to {Rx1}. 28.004 DISP Short the wires from the UUT together at 28.004 DISP the Test Interface Panel. 28.004 DISP Adjust the ZERO OHMS control for 0 reading on the meter. 28.004 DISP Remove the short. # Low Ohms readings taken by HP 3458A 28.005 DISP The HP 3458A will be used in 4-Wire Ohm Mode to 28.005 DISP measure the actual resistance of the 5100B. 28.006 DISP [32]Connect the 5100B to the 3458A as follows: 28.006 DISP 28.006 DISP 5100B TO 3458A 28.006 DISP OUTPUT HI ---------> INPUT HI 28.006 DISP SENSE HI ---------> SENSE HI 28.006 DISP OUTPUT LO ---------> INPUT LO 28.006 DISP SENSE LO ---------> SENSE LO 28.007 RESF 10.0Z S 4W #! System Uncertainty not calculated because tolerance or floor unspecified. 28.008 3458 10.0Z E N 4W 28.009 DISP Disconnect the 5100B from the 3458A and 28.009 DISP connect the UUT to the 5100B as follows: 28.009 DISP 28.009 DISP 5100B TO VOM 28.009 DISP OUTPUT HI ÂÄÄÄÄÄÄÄÄÄÄ> Probe Tip 28.010 DISP SENSE HI Ù 28.011 DISP OUTPUT LO ÂÄÄÄÄÄÄÄÄÄÄ> Ground Clip 28.012 DISP SENSE LO Ù 28.013 MEME 28.014 MEMI Use keyboard to enter actual UUT reading in Ohms: 28.015 MEME 28.016 MEMC ÿê +8% -7.5% #! Test Tol 0.75, Sys Tol 0.0002, TUR 3750.00 (>= 4.00). 29.001 DISP Set the Range to {Rx100}. 29.002 DISP Short the wires from the UUT together at 29.002 DISP the Test Interface Panel. 29.002 DISP Adjust the ZERO OHMS control for 0 reading on the meter. 29.002 DISP Remove the short. 29.003 RESF 1000Z +8% -7.5% 4W #! T.U.R. not calculated because tolerance or floor unspecified. 30.001 DISP Set the Range to {Rx10,000}. 30.002 DISP Short the wires from the UUT together at 30.002 DISP the Test Interface Panel. 30.002 DISP Adjust the ZERO OHMS control for 0 reading on the meter. 30.002 DISP Remove the short. 30.003 RESF 100000Z +8% -7.5% 2W #! Test Tol 7500, Sys Tol 3, TUR 2500.00 (>= 4.00). #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 1 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.